15
TROUBLES AND POSSIBLE SOLUTIONS
Trouble Possible solutions
The equipment doesn’t turn on.
- Check if battery is well connected.
- Check if outlet is energized and in good condition.
- Recharge the battery.
- Call for DMC’s technical assistance.
The battery charge is exhausted quickly
- Check the battery charge level (see item “USING”) and recharge it. In
case of the charge level does not change upon charging, contact DMC.
Apparently weak laser emission
- Remove the spacer from the optical fiber and check for debris
accumulation on the emission side
- Replace the spacer.
DISCARD
The end of equipment life and its components may induce to environmental contamination or if they
are unduly used. To decrease these risks, the customer should discard the equipment according to
the local legislation.
The Therapy EC battery contains lithium and shall not be disposed in common waste.
LASER SUPPLY SYSTEM
The equipment is composed by two lasers, one red and one infrared. Laser light is conducted to the end
of handpiece by optical fiber. The luminous power emitted by laser diodes is controlled by electric current.
Hand Piece
Laser Diode
Optical
Fiber
Power
Supply
Control
Current
Monitor
Signal
Conditioner
Laser Output