5
GENERAL FEATURES
Features Specifications
Supply voltage 100-240 V~
Input current Alternating current
Output current Continuous current
Input power 25 VA
Working mode Continuous
Lasers working mode Continuous
Voltage frequency 50/60 Hz
Protection against shock Internally Powered
Protection degree regarding water and solid
object penetration
IP20
Uncertainty of time (seconds) ± 4%
Uncertainty of energy (Joules) ± 20.4%
Applied part Polymeric spacer
Handpiece size 3 cm (length) x 5 cm (width) x 21 cm (height)
Handpiece support size 6 cm (length) x 19 cm (width) x 11 cm (height)
Handpiece weight 0,18 Kg
Handpiece support weight 0,12 Kg
Optic fiber diameter 1000 µm each fiber
Battery Li-ion 3,7V 18650 with protection circuit
Manufactured and tested according to: IEC 60601-1, IEC 60601-1-2, IEC 62034 and IEC 60685-1.
Software version 11
Use temperature 10° - 30°C
SAFETY - IMPORTANT PRECAUTIONS
• Please read the whole manual before using the equipment.
• Attention - Using other controls, setting or running not specified in this manual may result in the
exposure to hazardous radiation.
• Attention - Laser smoke may present living tissue particles.
• As the equipment will be used by qualified professional, an additional specific training for its using
is not required. However, the reading of whole manual before its using is recommended.
• If the equipment is not used for some time, the battery shall be removed.