4 5
PRECAUTIONS
When to exercise special care when prescribing the IceMan®
Exercise special care prescribing this device for the following patients:
those with arthritic conditions; peripheral vascular disease; children
under the age of ; those with decreased skin sensitivity; poor circula-
tion, or compromised local circulation; hypercoagulation disorders;
diabetes or neuropathies.
WARNINGS & PRECAUTIONS (con’t)
• The barrier between your skin and the cold pad may develop
moisture during use, which may create colder temperatures on
the skin. Temperatures that are too cold may result in serious
injury, including tissue necrosis. ALWAYS check for moisture on
the barrier between your skin and the cold pad. If moisture is pres-
ent on the barrier, immediately discontinue use of this device.
• Poor connections between hoses may cause leaking, which may
result in serious injury, including infection and tissue necrosis. AL-
WAYS listen for a “snap” or “click” when connecting the IceMan®
cold therapy unit hose to the cold pad hose. Use only IceMan®
cold pads.
• Use of the IceMan® with wet hands or in a wet location may
result in electrical shock and serious injury. DO NOT handle
transformer or power cord with wet hands or in a wet location.
The power supply unit is the mains power disconnect. Do not
position the equipment to make access to the disconnect dif-
ficult. Only connect equipment to the power supply provided for
this product.
• DO NOT use the IceMan® near flammable anesthetics or oxygen
enriched environment, which may result in explosion and serious
injury.
• Keep power cord, hose, small parts, and packaging materials
away from children and animals. These items pose a risk for suf-
focation or strangulation.
• It could be unsafe to use accessories, detachable parts and
materials, or interconnect to other equipment not described in
these instructions, or otherwise modify the equipment.
• Care must be taken when operating this device adjacent to
other equipment. Potential electromagnetic or other interfer-
ence could occur to this or other equipment. Try to minimize this
interference by not using other electronic equipment in conjunc-
tion with this device.
• To avoid the risk of electrical shock, do not disassemble the
IceMan®. If device is not functioning properly, please contact
DonJoy product support.