Dornier Medilas H
Service Manual
Order number: K2011036 de
K2011037 en
WARNING
The Dornier Medilas H may only be serviced and maintenanced by authorized persons. Authorized
persons are persons trained by Dornier MedTech GmbH or by companies entrusted from Dornier
MedTech GmbH only.
Service and maintenance measures performed by unauthorized persons can cause life-threatening
injuries and/or serious damages on the Dornier Medilas H.

Table of Contents

Questions and Answers

  • A
    Alec FosterAug 2, 2025
    What to do if no laser pulse is emitted from Dornier Medical Equipment?
    • B
      Beth BennettAug 2, 2025
      If your Dornier Medical Equipment isn't emitting a laser pulse, consider these potential causes and solutions: * The foot switch might be defective and needs replacement. * The air break switch may be defective; in this case, replace the air break switch. * Check for any leaks in the tubing, and replace the tubing if necessary. * A defective CPU could also be the reason; consider replacing the CPU board. * Inspect Cable W4; ensure it is properly connected, and replace or repair it if damaged.
  • J
    johnsonrachelAug 6, 2025
    Why does my Dornier Medical Equipment keep reporting overtemperature (error code "1")?
    • S
      Samuel BurnettAug 7, 2025
      If your Dornier Medical Equipment is repeatedly reporting an overtemperature error, here are some potential causes and solutions: * The location might have restricted air change; try changing the location. * The cooling system may be dusty and needs cleaning. * The cooling fan may not be running; ensure it is properly connected or replace it. * The ambient temperature could be too high; provide sufficient room cooling or change the location. * The temperature sensor in the cooling system may need calibration. * Cable W13 might be faulty and needs replacement or repair. * The CPU could be defective, requiring replacement of the CPU board.
  • P
    Phyllis EllisAug 12, 2025
    How to fix a pilot laser that is not visible on Dornier Medical Equipment?
    • C
      Caitlin RodriguezAug 12, 2025
      If the pilot laser on your Dornier Medical Equipment is not clearly visible, consider the following: * The pilot laser might be set to level 0; try setting it to a higher level. * For serial numbers less than 79, the pilot driver (A10) may not be switched on; ensure it is switched on. * Check Cable W6 for proper connection; replace or repair if necessary. * The pilot laser might be calibrated incorrectly. * The fiber may not be adjusted correctly, or the optic may be defective; adjust the pilot laser coaxially to the therapy laser beam, adjust the fiber coupler, or replace the fiber. * The pilot laser or CPU could be defective.
  • B
    bowersaliciaAug 16, 2025
    What to do if the therapeutic effect is too low with Dornier Medical Equipment?
    • D
      Diana BowenAug 16, 2025
      If you're experiencing a low therapeutic effect with your Dornier Medical Equipment, here's what you can check: * The fiber or applicator might need replacement. * The fiber coupler may need adjustment; also, check the input lens and replace if necessary. * The resonator mirror may need cleaning or replacement. * The laser crystal may need replacing. * Calibration might be off; check the monitor adjustment and calibrate the device.
  • K
    Kevin WashingtonAug 22, 2025
    Why does the laser pulse abort prematurely on my Dornier Medical Equipment?
    • A
      Andrew PadillaAug 22, 2025
      If the laser pulse on your Dornier Medical Equipment is aborting prematurely, the cause might be: * A defective air break switch or pneumatic potentiometer; consider replacing them. * A faulty foot switch; try replacing it.
  • C
    Carla JohnsonAug 27, 2025
    What to do if the display is dark or key presses aren't recognized on Dornier Medical Equipment?
    • J
      Jason MooreAug 27, 2025
      If the display on your Dornier Medical Equipment remains dark, or key presses are not recognized, consider the following: * Cable W3 might be the issue; replace or repair it. * The control panel could be faulty; replace the control panel or let it dry slowly. * The CPU might be defective; consider replacing the CPU board.

Summary

Safety

Laser Safety

Details safety precautions regarding laser radiation, protective goggles, and laser area management.

Protective Goggles

Specifies requirements and inscriptions for safety goggles used with the Dornier Medilas H.

Installation

Installation; first start-up

Details procedures for initial setup, including mains voltage adjustment and first operation.

Description

Laser safety chain

Describes the integrated safety chain that prevents unintentional laser emission.

Service mode

Instructions on how to access and use the service mode for diagnostics and settings.

Calibration

Procedures for calibrating various aspects of the Dornier Medilas H.

Checking, replacement, adjustment

CPU board

Procedures for replacing the CPU board and its components.

Replacing laser power supply

Guide to replacing the laser power supply, including detaching connections and mounting brackets.

Adjusting resonator

Procedures for adjusting the laser resonator for optimal beam output.

Maintenance

CPU - Software Upload

Steps for uploading new software to the CPU via a PC.

Troubleshooting

Errors with displayed error codes

Lists error codes, causes, checks, and measures for troubleshooting.

Errors without displayed error codes

Troubleshooting for issues not accompanied by error codes.

Dornier Medilas H Specifications

General IconGeneral
BrandDornier
ModelMedilas H
CategoryMedical Equipment
LanguageEnglish