Display-indication with active calibration code(OP1):
The CAL -segment and the small 1 on the left indica-
tes to the user that oP1 is activated.
Calibration by physical standard references / oP2
Use the up and down keys [

] to select oP2. Push
[ENTER/MENUE] to confirm.
On the bottom of the display appears 1 P. For chang-
ing between a 1-Point [1 P], 2-Point [2 P] or 3-Point
[3 P] - calibration use the up and down keys [

].
Example of a 1-Point calibration:
Push [ENTER/MENUE] to confirm. On the display
appears Go. After the displayed measuring value is
stabile push [ENTER/MENUE]. App. 2 seconds later
in the first display line appears P1 (measuring value
1); in the second display line appears dP. (standing
for decimal point).
Use the up and down keys [

] to select the num-
ber of digits after decimal point:
dP. = two decimal digits
dP . = one decimal digit
(decimal point is jumping one digit to the right)
Push [ENTER/MENUE] to confirm. On the second
line of the display appears Si -. Use the up and down
keys [

] to select the requested sign.
Si = reference standard shows
a negative measuring value (below 0,00C°)
Si = reference standard shows
a positive measuring value (above 0,00C°)
Push [ENTER/MENUE] to confirm. On the second
line of the display appears Fd 0.
Use the up and down keys [

] to select the re-
quested range:
Fd 0 = below 100,00°C
Fd 1 = above 100,00°C
Push [ENTER/MENUE] to confirm. On the display
appears 00.00. For changing the value use the up
key [
] . For stepping to the next number use the
down key [
]. If the requested value is complete
then push [ENTER/MENUE] to confirm.
Push [ESC] to be back in the measuring mode.
Note: A break of the physical calibration cannot
be done by the [ESC]-button. Switching off the in-
strument can only do a break of the physical cali-
bration.

oP2
1 P 2 P 3 P
Important!
9
DE P700 Anleitung engl. 12/11 06.12.2011 10:04 Uhr Seite 9

Other manuals for Dostmann Electronic P700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dostmann Electronic P700 Series and is the answer not in the manual?

Dostmann Electronic P700 Series Specifications

General IconGeneral
BrandDostmann Electronic
ModelP700 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals