Remote Operation
Example
“SYST:ERR?”
Result was “1, INVALID CHARACTER”, check for more errors.
Description: This error is no longer supported. The error code is
maintained and reserved for legacy purposes only.
“SYST:ERR?”
Result was “2, OUTPUT BUFFER OVERFLOW”, check for more errors.
Description: This error is no longer supported. The error code is
maintained and reserved for legacy purposes only.
“SYST:ERR?”
Result was “3, TOO MANY COMMANDS”, check for more errors.
Description: The maximum of 220 characters per command line has been
exceeded.
SYST:ERR?”
Result was “4, SYNTAX ERROR”, check for more errors.
Description: There is a misspelling in your command or a non-numeric
character was included in a command where a number should have been, or
use of unrecognized symbols such as %, &, #, etc.
SYST:ERR?”
Result was “5, DATA OUT OF RANGE”, check for more errors.
Description: The value transmitted is not acceptable.
Eg. 1: A non existing switch positions has been commanded. Sending
Route:Switch1 8. When switch 1 is a SP6T (6 position switch)
Eg. 2: This error code is set if the IP address or MAC address is an invalid
one. Sending SYSTEM:IPADDRESS 55.57.2 would generate this error
code since 4 numbers are required for a valid IP address.
SYST:ERR?”
Result was “6, ILLEGAL PARAMETER VALUE”, check for more errors.
Description: This error is no longer supported. The error code is
maintained and reserved for legacy purposes only.
SYST:ERR?”
Result was “7, INPUT BUFFER UNDERFLOW”, check for more errors.
Description: This error is no longer supported. The error code is
maintained and reserved for legacy purposes only.
SYST:ERR?”
Result was “8, MATRIX SOCKET NOT AVAIL”, check for more errors.
Description: This error is no longer supported. The error code is
maintained and reserved for legacy purposes only.