5.
M
AINTENANCE
I
NFORMATION
Daily: Check the filter glass for damage and dirt; check the housing for damage.
Every 400 operation hours / each 2-3 years in private operation:
Replace the lamps.
Every 2 years: Maintenance by our Customer Service or a technician authorized by us
(applies only to commercial users but is recommended to comply with the
Medical Devices Act).
5.1. C
LEANING THE
H
AND
L
AMP
Disconnect the unit by unplugging the power-pack!
Never wipe the hand lamp or the filter glass with a dry cloth. To clean dusty or dirty surfaces,
wipe with a damp soft cloth or sponge (add a little dishwashing liquid to warm water).
Anti-static plastic cleaners are also suitable.
Never clean with concentrated disinfectant or solvent such as Sagrotan, Lysoform,
methyl alcohol or any other liquid containing alcohol.
5.2. C
LEANING AND
D
ISINFECTING THE
C
OMB
A
TTACHMENT
The comb attachment (Art.-No. 995200) is skin-friendly and hygienic. Its poreless surface
makes it difficult for dirt and bacteria to accumulate.
To clean and disinfect the comb, see Section 5.2.
• At 25 °C the comb is resistant to the following chemicals:
- Baktolan, up to 5 % - Merfen
- Chinosol, up to 1 % - Perhydrol
- Elmocid Gamma, up to 2 % - Perodin
- Lugol's solution - Sagrotan, up to 2 %
- Mefarol, up to 1 % - Hydrogen peroxide, up to 40 %
- Merckojod, up to 1 % - Zephirol, up to 5%