Dräger FPS-COM 5000 25
Troubleshooting
Mechanics
Battery compartment
cannot be closed/
opened
Screw not completely
loosened
Completely loosen
the screw.
Communication unit
cannot be fitted/
dismantled
No fixing lugs/
defective fixing lugs
on visor frame of
mask
Replace visor frame.
Fastening eyes on
communication unit
faulty
Replace fastening
eyes.
Communication unit
not properly hooked
onto connector
Dismantle
communication unit
and fit again.
Earphone cannot be
bent to desired
position
Earphone deformed
or gooseneck faulty
Replace earphone.
Fault Cause Remedy
No connection via
cable to radio
possible
Cable faulty Replace cable.
Cable not fitted
correctly
Check connection
between cable and
radio. Fit cable again
if necessary.
Check connection
between cable and
communication unit.
Fit cable again if
necessary.
Incorrect cable
connected
Use suitable cable.
Fault Cause Remedy