Technical Data
Ambient conditions
During operation:
Temperature 15 °C to 30 °C
Atmospheric pressure 900 hPa to 1100 hPa
Relative humidity 0 to 90 %, no condensation
Air velocity Max. 0.3 m/s
During storage/transport:
Temperature –20 °C to +60 °C
Atmospheric pressure 900 hPa to 1100 hPa
Relative humidity 0 to 90 %, no condensation
Radiant heater
Radiant power
Light Halogen, 12 V / 35 W
Timer 0 to 99 minutes 59 seconds;
Start and stop times programmed in seconds;
Reset function;
Audible signal at 1 minute, 5 minutes and 10 minutes
Clearance between top edge of
radiant heater housing and ceiling >50 cm
53
Technical Data
Distance between radiant
heater and resting surface
of bed, in cm
Radiated power at
heater level 5,
in mW/cm
2
Radiated power at
heater level 10,
in mW/cm
2
80 7 14
70 9 18
60 10 25