EasyManuals Logo

Durr Dental VistaIntra User Manual

Durr Dental VistaIntra
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
Error Possible cause Remedy
X-ray image too bright
Detector being used not com-
patible with the unit settings
Wait until the x-ray emitter has
cooled down.
X-ray emitter incorrectly posi-
tioned
Correct the position of the X-
ray emitter
X-ray parameters set incorrectly
Check X-ray parameters and
adjust if necessary. Increase
the exposure time.
Inform a service technician.
Detector not correctly positioned
in the mouth of the patient
Correct the position of the
detector
X-ray image too dark
Detector being used not com-
patible with the unit settings
Use a different detector or
change the unit settings.
X-ray parameters set incorrectly
Check X-ray parameters and
adjust if necessary. Reduce
the exposure time.
Inform a service technician.
Troubleshooting
30 2202100028L02 2007V019
EN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental VistaIntra and is the answer not in the manual?

Durr Dental VistaIntra Specifications

General IconGeneral
BrandDurr Dental
ModelVistaIntra
CategoryMedical Equipment
LanguageEnglish

Related product manuals