EasyManua.ls Logo

DV Power TRT3 Series - Technical Data; Mains Power Supply; Output Data; Measurement

DV Power TRT3 Series
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TRT3x Technical Data
M-T03XNN-205-EN 18
8 Technical Data
8.1 Mains Power Supply
- Connection: According to IEC/EN60320-1; UL498, CSA 22.2
- Voltage: 90 264 V AC, or 110 350 V DC
- Frequency: 50 60 Hz
- Input power: 200 VA
- Fuse: 2 A / 250 V, type F, but not user replaceable
8.2 Output data
Model
Output voltages available
Model
Output voltages available
TRT30A
8, 40, 100 V AC
3 x (8, 40, 100)√3 V AC
TRT33A
1, 8, 40, 100 V AC
3 x (1, 8, 40, 100)√3 V AC
TRT30B
10, 40, 100 V AC
3 x (10, 40, 100)√3 V AC
TRT33B
1, 10, 40, 100 V AC
3 x (1, 10, 40, 100)√3 V AC
TRT30C
8, 40, 80 V AC
3 x (8, 40, 80)√3 V AC
TRT33C
1, 8, 40, 80 V AC
3 x (1, 8, 40, 80)√3 V AC
8.3 Measurement
- Turns ratio measuring range: 0.8 to 15 000
- Turns ratio resolution: 5 digits
- Typical turns ratio accuracy:
0.8 - 999: ±0.05 %
1000 - 3999: ±0.05 %
4000 - 15000: ±0.1 %
@80 or 100 V AC
0.8 - 999: ±0.05 %
1000 - 3999: ±0.1 %
4000 - 15000: ±0.2 %
@40 V AC
0.8 - 999: ±0.05 %
1000 - 3999: ±0.1 %
4000 - 15000: ±0.2 %
@8 or 10 V AC
0.8 - 999: ±0.05 %
1000 - 1999: ±0.1 %
@1 V AC
- Guaranteed turns ratio accuracy:
0.8 - 999: ±0.125 %
1000 - 3999: ±0.125 %
4000 - 15000: ±0.25 %
@80 or 100 V AC
0.8 - 999: ±0.125 %
1000 - 3999: ±0.25 %
4000 - 15000: ±0.5 %
@40 V AC
0.8 - 999: ±0.125 %
1000 - 3999: ±0.25 %
4000 - 15000: ±0.5 %
@8 or 10 V AC
0.8 - 999: ±0.125 %
1000 - 1999: ±0.25 %
@1 V AC
- Excitation current range 0 2 A
- Typical excitation current accuracy ± (0.25% + 500 μA)
- Guaranteed excitation current accuracy ± (0.25% + 1 mA)
- Excitation current resolution
0 9.9999 mA 0.1 μA
10 99.999 mA 1 μA
100 999.99 mA 10 μA
1 2 A 100 μA
- Phase angle range 360 Degrees
- Typical phase angle accuracy ± 0.05 Degrees
- Guaranteed phase angle accuracy ± 0.1 Degrees
- Phase angle resolution 0.01 Degree