'LDJQRVWLFVDQG)DXOW)LQGLQJ
DX Power
Module
Status LED
Flash Code
Possible Cause and Action
60
11
Stall Timeout Fault
Cause: The motor current has been at, or close to, current limit
for longer than the Stall Timeout parameter value.
Action
< Turn the DX System off then on again. The system
may require time to cool down.
Cause: Motor(s) are faulty. Wheel(s) may be rubbing on frame.
Action
< Ensure wheels turn freely while under no load. Have
motor(s) checked by a service technician.
12
Module Mismatch
Cause: The data held by the DX Remote for the DX Power
Module is corrupt, incompatible or non-existent.
Action
< Reprogramming the wheelchair system may correct
this problem.
Cause: There is a compatibility problem between DX Modules
in the System. The wheelchair will be disabled.
Action
< Consult your Dynamic Service Centre.