13 Dynatron® ibox™ Operator’s Manual
|
Rev. 2
|
5/14/2019The Dynatron® ibox™ Iontophoresis Device
table of contents symbols & labels limited warrantyprev
< >
nextprev topic
<
< >>
next topic
Solutions: 1) Avoid abrupt movements of the electrodes or patient
during a treatment.
2) Avoid contact with any external electrical sources.
3) If this message occurs during a treatment, consider
setting-up a similar treatment, but use a dose equivalent
to the remaining dose that would have been delivered
if the device had not shut-down due to the over-current
condition.
“MSP430 ERROR” - This error message is displayed if the internal mi-
crocontroller circuit encounters difculty controlling the device. If this
condition occurs, any treatments in progress will be discontinued and the
outputs turned-off. Although this is a rare condition, if it is recurrent, the
user should not use the device for any further treatments and Dynatronics
should be contacted for servicing of the device.
Solutions: 1) If not a recurrent error, consider continued use of
the device by turning it off and then on again.
2) If a recurrent error, do not use the device for treatments.
Contact Dynatronics for servicing of the device -
(800) 874-6251.