Technical
Information
The following provides technical information and guidelines for the
maintenance and care of the Dynatron ibox.
Outputs
Output Jacks: 1
Current (Intensity): 0.5 – 4.0 mA
Current Accuracy: +/- 10%
Current Precision: 0.1 mA
Voltage at Electrodes: +/- 0-46 Volts
Electrode Cable Jack: Custom 2-pin polarized
Treatment Timing
Maximum Total Deliverable Dose: 3360 mA-min.
(using high quality Alkaline batteries)
Single Treatment Deliverable Dose: 1-160 mA-min.
Number of 40 mA-min. Treatments: 89
Minimum Treatment Time: 15 sec.
Maximum Treatment Time: 100 minutes
Timing Accuracy: +/- 1 second
Dimensions
Length: 4.5”
Width: 2.66”
Height: 1.25”
Weight: 3.6 oz. (without battery or electrode cable)
Display: Organic Light Emitting Diodes (OLED)
Environmental Conditions
Transport and Storage
a. Ambient temperature range: -10C
+50C
b. Relative humidity: 10% to 100% non-condensing
c. Atmospheric pressure: 200 hPa to 1060 hPa
Operation
a. Ambient temperature range: -10C
+40C
b. Relative humidity range: 30% to 75% non-condensing
c. Atmospheric pressure range: 700 hPa to 1060 hPa
If it is necessary to store the Dynatron ibox for an extended period
(one month or more), remove the batteries and store the device in a
cool, dry location.