TESTING
Note: Make sure the solution is at room temperature 59°F to 95°F (15°C to 35°C) prior to testing.
1. Insert a Test Strip, Contact Bar's end rst, into the test port. The Meter will automatically turn on. A beep will sound and moving
dotted lines will appear followed by a ashing “ “ symbol.
2. After the “ “ symbol appears, press the S button for 3 sec to switch control solution mode. The “ctl“ symbol indicates that the Meter
will mark your current test as a Control Solution test. Now you are ready to apply the Control Solution.
3. Discard the rst drop of Control Solution and squeeze a small drop of Control Solution on a clean nonabsorbent surface
(such as a clean piece of wax paper). Do not apply Control Solution to the Test Strip directly from the bottle.
4. Bring the tip of the Test Strip to lightly touch the drop of Control Solution. Control Solution is automatically pulled into the strip
through the tip. Hold until the meter beeps. The meter will now being counting down from 5 to 1 and the Control Solution test
result will appear.
5. Compare the results with the Control Solution range printed on the Test Strip vial. The results should fall within the specied ranges
listed on the Test Strip vial.
COMPARING CONTROL SOLUTION RESULTS
If the test result falls outside the specied range printed on the Test Strip vial, repeat the test.
RESULTS FALLING OUTSIDE THE SPECIFIED RANGE MAY BE CAUSED BY:
●
Error in performing the test
●
Control solution temperature is lower than 59°F (15°C) or higher than 95°F (35°C)
●
Expired or contaminated Control Solution
●
Expired or contaminated Test Strips
●
Meter malfunction
If the problem persists, the system may not be working properly. DO NOT continue to use the system to test your blood glucose level.
If you continue to get results that fall outside the specied range, please call MHC Medical Products, LLC customer service line at
1.877.358.4342 or e-mail us at info@MHCmed.com
STORAGE AND STABILITY
For best results, store Control Solution at room temperature; do not refrigerate. Refer to the expiration date for stability.
The solution should NOT be used after the expiration date or 3 months after opening. For in vitro diagnostic use only.
Made in Taiwan. Distributed by:
..
CAUTION:
•
Use the same model of Control Solution, Test Strip and Glucose Meter.
•
DO NOT apply Control Solution to the Test Strip directly from the bottle, such operation could be leaded to contaminate
and decrease the concentration of Control Solution. You MUST place a drop of Control Solution on a clean nonabsorbent
surface rst.