Loading...
The
selection
of
the
"BASE"
units
is
now
possible.
The
choices
available
for
base
radiation
units
are:
R
roentgen
cnt
counts
Gy
gray
Sv
sieverts
rem
roentgen
equivalent
man
dis
disintegrations
rad
rad
Notice
that
the
base
unit
is
what
is
displayed;
thus,
if
mR/h
is
desired
on
the
display,
select
"К"
as
the
base
unit.
The
prefix
(milli)
and
the
suffix
(h)
will
be
added
in
the
next
steps.
In
each
case,
press
-
to
reject
the
displayed
parameter,
and
another
selection
will
be
offered.
When
the
display
shows
the
desired
base
unit,
press
+
to
accept
it.
CAUTION
Selection
of
a
new
base
unit
requires
an
ap-
propriate
change
in
the
calibration
constant
(CC).
Refer
to
section
ILE.3.
Changing
the
prefix
or
suffix
does
not
affect
calibration
and
so
requires
no
change
in
the
calibration
constant.
Selection
of
"SUFFIX"
After
the
4
is
pressed
(as
discussed
above)
to
accept
the
base
unit,
the
display
will
then
present
the
selection
of
suffix
as
follows:
SUFFIX
(UNIT
TIME)
+
=
USE/-
=
NO
The
suffix
is
the
unit
of
time
used
to
calculate
the
displayed
"RATE
METER
READING."
Three
are
available:
$
second
min
minute
h
hour
Again,
press
the
-
switch
to
reject
the
suffix
displayed
and
call
the
next
choice
to
be
displayed.
When
the
desired
suffix
is
displayed,
press
+
to
accept
it.
Selection
of
"PREFIX"
2-10
ESPI.MAN/February
1992
~

Related product manuals