Model
HP-190-A
Hand
Probe
A.
GENERAL
DESCRIPTION
The
Model
HP-190A
hand
probe
uses
a
thin
end-
window
GM
tube
for
detection
of
relatively
low
energy
beta
and
provides
a
limited
sensitivity
to
high
energy
alpha
particles.
The
HP-190A
does
not
have
energy
compensation;
therefore,
it
is
not
recommended
for
gamma
exposure
rate
measurements.
B.
SPECIFICATIONS
Operating
Voltage:
900
+
50У
Plateau
Length:
100
Y
minimum
Plateau
Slope:
0.1
percent
per
V
maximum
Dead
Time:
200
zs
maximum
Temperature
Range:
—67?
F
to
+
167°
F
(—55°
C
to
+75°
C)
Mica
Window
Thickness:
1
!/8-
inch-diameter
(2.9
cm)
THIN
END
WINDOW
HIGH
BETA
SENSITIVITY
WINDOW
PROTECTIVE
SCREEN
Gamma
Sensitivity:
z
2500
cpm/mR/h
(9105)
(into
window)
*Beta
Efficiency:
=35
percent
*Sr-°°Y
2:25
percent
°Tc
=
10
percent
"C
Alpha
Sensitivity:
3
MeV
at
window
Connector:
BNC
series
coaxial
Size:
1
2/8
inches
in
diameter
x
4
7/8
inches
long
(3.5
cm
x
12.4
cm)
Weight:
5.5
ounces
(155
g)
*Measured
without
screen
cap.
With
cap
in
place,
effi-
ciency
will
be
Ξ:50
percent
of
values
listed.
Efficiencies
listed
as
percent
of
27
emission
rate
from
a
one-inch-
diameter
source.