5.2.7.5.6 The result of the dosimeter verification is considered positive
if the main relative permissible error limit of DE measurement at 0.95
confidence probability does not exceed 15 %.
5.2.7.5.7 If the limit of main relative permissible error of DE
measurement does not meet the requirements stated in 5.2.7.5.6, the
dosimeter cannot be verified and should be sent for repair.
5.2.7.6 Presentation of verification results
5.2.7.6.1 Positive results of periodic verification and verification after
repair are registered in the Table of Appendix G, or by issuing a certificate
of verification of legally regulated measuring equipment.
5.2.7.6.2 If the dosimeter is recognized inapplicable for use as a result of
verification, it gets the certificate of inadequacy.