EasyManua.ls Logo

ECOTEST DKG-21 - Page 9

ECOTEST DKG-21
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
3.2.9 Time of operating mode setting at DER measurement, not more
than:
- from 1∙10
-6
Sv/h to 510
-6
Sv/h (inclusive), maximum 30 minutes;
- from 5∙10
-6
Sv/h to 210
-5
Sv/h (inclusive), maximum 5 minutes;
- from 2∙10
-5
Sv/h to 1.0 Sv/h, maximum 3 min.
3.2.10 Time of DER measurement, not more than:
- 10 s in the DER range of 1.0 μSv/h to 10.0 μSv/h;
- 5 s in the DER range of 10.0 μSv/h to 1.0 Sv/h.
3.2.11 Minimum DE measurement time expressed in seconds should
not exceed the value calculated by the formula:
, (1.1)
where
)10(
p
H
- a numeric value of DER in µSv/h.
)10(
7200
p
H
t
=

Related product manuals