For single use only
These are general instructions for setting up a pressure monitoring
system and/or Edwards compatible hardware with the FloTrac
sensor. Since kit configurations and procedures vary according to
hospital preferences, it is the responsibility of the hospital to
determine exact policies and procedures.
Caution: The use of lipids with the FloTrac sensor may compromise
product integrity.
Concept/Description
The FloTrac sensor is a sterile, single use kit that monitors pressures
when attached to pressure monitoring catheters. The disposable sterile
cable with a white connector interfaces exclusively with an Edwards
cable that is specifically wired for the pressure monitor being used. The
disposable sterile cable with a green connector interfaces exclusively
with the Edwards APCO cables for use with the Edwards arterial
pressure based cardiac output monitoring devices or hardware.
The FloTrac sensor has a straight, flow-through design across the
pressure sensors with an integral flush device.
Indications
The FloTrac sensor is indicated for use in intravascular pressure
monitoring. It is also indicated for use with the Edwards arterial pressure
based cardiac output monitoring devices or hardware to measure
cardiac output.
FloTrac Sensor
Edwards Lifesciences, the stylized E logo, Edwards, Vigileo and
FloTrac are trademarks of Edwards Lifesciences Corporation.
1
T4066
FloTrac Sensor
Sensor Vent Port
Snap-Tab Device
To Vigileo monitor (green)
To
IV Set
To Patient
To Pressure monitor (white)
DIRECTORY
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Français . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Deutsch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Español . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Italiano . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Nederlands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Dansk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Svenska . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
∂ÏÏËÓÈο . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Português do Brasil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
âesky
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
magyar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Polski . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Question and Answer IconNeed help?

Do you have a question about the Edwards Lifesciences FloTrac Sensor and is the answer not in the manual?

Edwards Lifesciences FloTrac Sensor Specifications

General IconGeneral
CategoryAccessories
SterilitySterile
Single UseYes
TypeSensor
MeasurementContinuous cardiac output
Parameters MeasuredCardiac output, Stroke volume, Systemic vascular resistance
TechnologyArterial pressure waveform analysis
CalibrationNot required
UseHemodynamic monitoring
ApplicationCritical care, operating room

Summary

Directory

Concept/Description

Indications

Contraindications

Procedure

Zeroing and Calibration

Testing Dynamic Response

Describes how to test the system's dynamic response using an oscilloscope or monitor.

Routine Maintenance

Warnings

Complications

Sepsis/Infection

Discusses risks of infection from contamination of the pressure setup.

Air Emboli

Explains how air can enter the patient's system through stopcocks or disconnections.

Clotted Catheter and Bleed-Back

Describes risks of blood bleed-back and catheter clotting if flush system is under-pressurized.

Overinfusion

Warns about potential harm from excessive flow rates exceeding 300 mmHg.

Abnormal Pressure Readings

Covers causes of abnormal readings like calibration loss, loose connections, or air.

How Supplied

Technical Assistance

Specifications

References

Symbol Legend