WARNING: UNEXPECTED RESTART
Failure or malfuncon of the central control system (unexpected start-up) or danger of a
pump restarng automacally aer switching o due to a fault.
Risk of crushing, cung or cung o by touching moving parts due to recurring start
command or power supply.
Before carrying out maintenance or service work, disconnect the pump from the power
supply, secure it against being switched on again, determine that it is de-energised,
ground and short-circuited and cover / isolate adjacent live parts.
CAUTION: RISK OF INJURY
Pulling in of body parts and objects into the vacuum through open anges. Do not
operate the vacuum pump with open anges. When installing the vacuum pump, rst
mechanically connect the inlets and outlets and then make the electrical connecons.
CAUTION: HIGH PRESSURE GAS
Danger due to incorrect connecon of the pump to the system to be pumped out.
Pressure build-up in the vacuum system.
Check correct connecon of inlet and outlet anges before commissioning. When using
the pump with seal gas, protect the seal gas supply so that no overpressure can occur in
the pump system in the event of a malfuncon or power interrupon.
CAUTION: TRIP HAZARD
Slipping, tripping or falling as a result of oil leaking from the pump. Oil can escape from
the vacuum pump during pumping or depending on the work process. Risk of falling on
oil spills. Check the vacuum pump regularly for oil leaks. Take the appropriate safety
measures
2.2 Electrical hazards
DANGER: ELECTRICAL SHOCK
Risk of electrical shock caused by direct or indirect contact with live parts. Electrical
shock due to faulty electrical connecon and possible residual voltage up to 5 minutes
aer disconnecon from the mains. The electrical connecon may only be carried out by
a trained person. Observe the naonal regulaons in the user country, e.g. for Europe EN
50110 - 1. Prior to servicing, disconnect the vacuum pump from the power supply.
Risk of electrical shock due to interrupon of the protecve conductor system. In the
event of a fault, life-threatening voltages may be present on electrically conducve
components. Before commissioning, check the resistance of the earthing cable and the
suitability of the assigned overcurrent protecon device.
Page 9
A41802880_C - Important safety informaon