Why are measurement values fluctuating on my Elitech ChloroChek Chloridometer 3400?
- PphawkinsJul 25, 2025
Fluctuating measurement values during measurement of standard solution or controls with Elitech Test Equipment can occur due to contaminated electrodes. Clean the electrodes with a silver cleaning cloth and rinse with distilled water. It can also happen if the electrodes are mechanically defective; in this case, replace the electrodes and working solution.