Questions and Answers

  • R
    Robert TuckerAug 11, 2025
    What to do if front panel LCD display or indicators do not come on for EME Medical Equipment?
    • A
      Andrew SalinasAug 11, 2025
      If the front panel LCD display or indicators don't come on and the unit doesn't work, first ensure the mains plug is properly plugged in and the mains socket is working. Then, check that the mains cable is correctly connected to the equipment. If the issue persists, replace the mains cable if it's worn or blocked. Also, verify that the power switch is on and replace any defective or blown fuses. If none of these steps resolve the issue, it may indicate an electronic control circuit malfunction.
  • C
    Crystal BowmanAug 20, 2025
    What to do if EME Shock Med Medical Equipment device does not emit shockwaves?
    • C
      Crystal VillarrealAug 20, 2025
      If your EME device isn't emitting shockwaves, start by carefully checking the output connections for correctness and integrity. If the connections are fine, inspect the applicator cable for wear or damage, and replace it if necessary. It could be that the applicator cable is interrupted or wrongly connected.
  • M
    Mrs. Susan Smith DDSAug 30, 2025
    Why is there a fall in treatment effectiveness with my EME Medical Equipment?
    • A
      alexandramayAug 30, 2025
      A fall in treatment effectiveness may be due to several reasons. First, ensure the output circuit of the applicator is efficiently connected. Perform regular maintenance, install the device as suggested, and check the integrity of the applicator's cable and connector. The actuator of the applicator might be damaged or worn and require replacement. Also, check the dispensing head (or transmitter) for mechanical damages, ensuring it maintains perfect cohesiveness with the treatment surface.

EME Shock Med Specifications

General IconGeneral
CategoryMedical Equipment
Frequency1-21 Hz
Pre-set ProtocolsYes
PressureUp to 5 bar
WaveformRadial