Quick Start Instructions
Important: Be sure your Empi Active device is turned off (in this condition, the light is
off for more than 5 seconds) before connecting it to an electrode or specialized wrap.
1. Wash and dry the skin area where you will apply the electrode or specialized wrap.
2. Connect a charged Empi Active device to an electrode or specialized wrap.
3. Apply the electrode or specialized wrap (with electrodes applied) to your skin.
4. Press
The light will glow blue steadily for about five seconds.
5. Press until you feel the desired intensity.
The light will flash blue every 5 seconds to indicate that the device is operating.
Change Intensity
When your Empi Active device has been on for 20 seconds and no buttons have
been pressed, the unit automatically becomes “locked” at the set intensity. Press
to unlock the unit. Then press either or to adjust the intensity of
your therapy.
If you press
when the unit is already at its lowest possible setting — or press
when the unit is already at its highest possible setting — the blue light will flash
quickly.
End a Therapy Session
1. To end a therapy session, press and hold for at least 1 second.
The light will first glow steady blue for a few seconds and will then flash four times.
This is the “goodbye” signal. Your Empi Active device is now turned off.
2. Carefully remove the electrode or specialized wrap from your skin. (Leave
the device attached to the specialized wrap unless you need to recharge
the device.)
3. Follow the instructions on the electrode or specialized wrap package
for storage.
Empi, Inc.
205 Highway 22 East
Clear Lake, SD 57226
800.328.2536
www.empi.com
360372-001 Rev. B; © 2009, 2011 Empi 05/11 DJOglobal.com
13051_360372A.indd 24 6/7/11 8:27:03 PM