EasyManua.ls Logo

Endress+Hauser FQG61 Series - Safety instructions: Special conditions; Safety instructions: Installation

Endress+Hauser FQG61 Series
20 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EG 10 003 / XA00507F-B Radiation Source Container
10 Endress+Hauser
Safety instructions:
Special conditions
Avoid electrostatic charging of coated and plastic surfaces. Do not rub.
Do not use in media or environments which may generate electrostatic charges on the coated and plastic
surfaces.
Avoid impact and friction sparks.
Safety instructions:
Installation
The suitability of the radiometric measurement method and of the device for applications in explosion
hazardous areas has to be checked by the operator of the plant according to national regulations.
Comply with the installation and safety instructions in the Operating Instructions.
Install the device according to the manufacturer's instructions and any other valid standards and regulations.
Do not operate the device outside the specified electrical, thermal and mechanical parameters.
Do not operate the device in environmental conditions that can result in the corrosion of the device or in
the pneumatic actuator.
Connect the apparatus to the on-site potential equalisation line.
Do not disconnect the compressed air supply in explosive atmospheres.
FQG_01
å 1
1 Potential equalisation
2 Compressed air supply
3 Certified associated apparatus
4 Proximity switches Ex ia
Radiation Source Container,
Type
Category Ambient temperature range Gas group Temperature class
FQG61-..L...........
FQG61-..N...........
FQG62-..L...........
FQG62-..N...........
II 2 G –20 °C Ta +80 °C IIB T4
1
3
3
2
4
A