SCOPE OF SUPPLY
For accurate measurement it is essential that the temperature of the probe and mainly of the sensing head is same as the
temperature of the air to measure.
Avoid mounting the EE150 sensor in a way which creates temperature gradients along the probe. If possible, EE150 shall be
installed with the entire probe inside the duct. For installation with mounting flange, in case of different temperature inside and
outside the duct, the probe part outside the duct shall be thermally isolated.
The device and mainly the sensing head shall not be exposed to extreme mechanical stress.
Do not attempt to remove the filer cap, which is fixed. Avoid touching the sensing head at all times.
The stainless steel probe is ESD sensitive and shall be handled as such. Do not connect it to the ground potential.
CAUTION
USER Manual
CONNECTION DIAGRAM
EE150 - Humidity and Temperature Sensor for HVAC applications
(class III)
(class III)
setup and adjustment interface (with EE-PCA)
setup and adjustment interface (with EE-PCA)
EE150 sensor, type number according to order (for ordering guide see data sheet at www.epluse.com/ee150)
Cable gland M16 x 1.5
Mounting flange, PC, Ø 6.0mm (with Type T2)
Test report according to DIN EN 10204-2.2
EE150-M1A6xx
EE150-M1A3xx
Find this document and further product information on our website at www.epluse.com/ee150.
DIMENSIONS / MOUNTING
Values in mm (inch)
CABLE GLAND M16x1.5
80.8
56
63
55.3
39.5
ø 5
ø 4
ø 6
ø 7
28.55
200
drilled hole for
mounting: ø>9
42
80.8 (3.18)
63 (2.48)
39.5 (1.56)
42
(1.65)
115.2
(4.54)
55.3
(2.18)
M12x1.5
M16x1.5
Duct mount
Wall mount

Other manuals for E+E Elektronik EE150

Question and Answer IconNeed help?

Do you have a question about the E+E Elektronik EE150 and is the answer not in the manual?

E+E Elektronik EE150 Specifications

General IconGeneral
Measurement ParametersRelative humidity and temperature
Humidity Range0 to 100 %RH
Protection ClassIP65
Output Signal4 to 20 mA, 0 to 10 V
Power Supply10 to 30 VDC
Operating Conditions0 to 100% RH
Weight200 g