Performing a Blood Glucose Test
Please adhere to your institution’s medical device handling and site preparation procedures.
1. Washyourpatient’shands
with soap and warm water.
Rinse and dry thoroughly. Your
institution may require that
you use an alcohol prep pad.
Please ensure that the test
site is completely dry before
proceeding.
2. Put a test strip into the meter
in the direction of the arrow.
The meter will automatically
turn on.
3. Please wait at least 5
seconds until the intended
area is dry and clean before
sampling the blood.
4. Whentheblooddropashes
on the display, obtain a drop
ofbloodfromngertip,using
the safety lancet. (See
image above)
5. Touch the blood drop to the
tip of the transparent window
of the test strip. Do not put
blood on top of the strip. Be
sure to get enough blood on
thestrip’sreactionzone.
Otherwise, an inaccurate
reading may result.
6. The meter will beep when
enough blood has entered
thestrip’sreactionzone.The
result will appear on the
display after 5 seconds.
7. Discard the used safety lancet
and test strip into an
appropriate sharps or
biohazardcontainer.
Page 9
EasyMax N System
Specications
Brand Name EasyMax N Blood Glucose Meter
Range 20~600mg/dL
Test Time 5 Seconds
Memory Sets 480 Test Results
Operating Condition
Temp. 50
0
F~104
0
F (10
0
C~40
0
C)
Relative Humidity R.H.
<
90%
Blood Sample
0.6 µL
Fresh Blood from Fingertip or Forearm
Hematocrit (Hct) 30~55%
Power 2 Alkaline 1.5V (AAA)
Battery Life Over 2,000 Tests
Display Dimension 1.60” x 1.30” (40.6 x 33.0 mm)
DeviceDimensionHxWxD(mm) 3.2” x 2” x 0.7” (81.3 x 50.89 x 17.8 mm)
Weight 1.6oz(45.5Grams)w/oBatteries
Principles Electrochemical Biosensor Technology
Page 10
In a Long Term Care facility setting, we recommend that a safety lancet be used for each
individual blood glucose test. For demonstration purposes, we are using SteriLance™’s
Pressure Activated Safety Lancets.
1. Carefully pull off the protective cap.
2. Pushthelancetrmlyontothechosensitetore.
3. Gently apply intermittent pressure near the puncture site to obtain the required blood volume.
EasyMax N system requires a small 0.6 microliter blood sample.