Operation, Installation and Calibration Instructions
N.I.S.T. Calibration Unit for Wrist Strap and Footwear Testers
Calibration Test Procedures
The calibration unit was specifically
designed for use in calibration of ESD
Systems.com brand test equipment.
The following step by step procedures
will cover calibration for specific test
units. The procedures will not cover
adjustment of the test equipment. For
detailed information regarding adjust-
ment of specific ESD Systems.com
brand testers contact our Customer
Service Department at (508) 485-7390.
CALIBRATION OF THE
MODEL 41193 WRIST STRAP
TESTER
Step 1: Connect one of the test leads
to common point banana jack labeled
"E" on 41230. Connect the opposite
end of test lead to one of the 41193’s
ground jacks located on the front edge
of the unit.
PS-2006 07/05 Page 1 of 2
Figure 1. 41230 Calibration Unit
Description
The ESD Systems.com model 41230
calibration unit is designed to simplify
the process of calibrating wrist strap
and footwear test equipment. The
calibration unit allows the user to
quickly and easily verify whether a
tester is operating within specifica-
tions. The 41230 is a passive device,
and requires no power source. The
calibration unit is manufactured within
industry accepted test ranges for both
wrist straps and footwear. The wrist
strap pass range is set at 750 Kilohm -
10 Megohm, while the footwear test
range is set at 750 Kilohm -
100 Megohm. The 41230 is supplied
calibrated to NIST traceable standards.
Installation
Remove the calibration unit from the
carton and inspect for shipping
damage. Each unit should include the
following:
1 - Calibration unit, item #41230
2 - 12'' test leads
Figure 2. Connecting test lead common
from 41230 to 41193.
Step 2: Connect the second test lead
to the 675 Kilohm jack labeled "A" on
41230. Touch the opposite end of test
lead to the test plate. Observe the
LED's for the proper response as
indicated below. Be sure to hold the
cord at an insulated point, so that the
resistance value is not affected by the
body.
Resistance Value Test Output - LED
(In Ohms)
Red jack “A” - 675K: Red (Low - hazard)
Green jack “B” - 825K: Green (OK - in limits)
Green jack “C” - 8.5M: Green (OK - in limits)
Yellow jack “D” - 11.5M: Red (High resistance)
Step 3: Repeat the procedure testing
across jacks B, C, and D. The different
resistance ranges should give the
display shown above. If the tester does
not meet calibration specifications,
contact factory for adjustment infor-
mation, or ask for Tech Brief PS-2055.
Figure 3. Testing 41193 for “PASS” and
“FAIL” ranges as indicated on
calibration unit.
Tech Brief
© 2005 DESCO INDUSTRIES INC.
Employee Owned
Question and Answer IconNeed help?

Do you have a question about the ESD SYSTEMS 41230 and is the answer not in the manual?

ESD SYSTEMS 41230 Specifications

General IconGeneral
BrandESD SYSTEMS
Model41230
CategoryTest Equipment
LanguageEnglish

Summary

Operation, Installation, and Calibration Instructions

Description of Calibration Unit

Explains the ESD Systems.com model 41230 calibration unit, its purpose, and specifications.

Installation of Calibration Unit

Instructions for unpacking and inspecting the 41230 calibration unit for shipping damage.

Calibration Test Procedures Overview

Introduces procedures for calibrating ESD Systems.com test equipment, not adjustments.

Calibration of Wrist Strap and Footwear Testers

Calibration of Model 41193 Wrist Strap Tester

Details steps for calibrating the Model 41193 wrist strap tester using the 41230 unit.

Calibration of Models 41201, 41202, 41203 Wrist Strap Testers

Describes the calibration process for 41201, 41202, and 41203 testers using the 41230 unit.

Foot Wear Test Circuit Calibration

Details the steps for calibrating the footwear test circuit using the 41230 unit.

Specifications and Warranty Information

Technical Specifications

Lists key specifications for the 41230 calibration unit including power, ranges, accuracy, weight, and dimensions.

Limited Warranty and Exclusions

Outlines the warranty period, coverage, exclusions, and liability limitations for the calibration unit.