5
LUBRICATION
Lubricate with Henry Schein Spray & Clean (102-8721) before sterilization.
1. Disconnect the handpiece from air supply.
2. Insert the nozzle into the handpiece drive air tube (Fig. 6) and spray for approximately
3 seconds.
3. Attach the handpiece to an air supply or purge station and run for approximately 20 seconds
to remove debris and excess lubricant. A spray cover or paper towel should be used to contain
expelled lubrication.
4. For push-button handpieces, it is recommended that the autochuck is periodically cleaned and
lubricated. Remove the bur, insert the spray nozzle tip into the chuck and spray for 1 - 2 seconds
(Fig. 7). Work the chuck mechanism to distribute the cleaner/lubricant by repeatedly pressing
the button. With the push-button depressed, insert and remove a bur several times to break-up
remaining debris.
Fig. 6
Fig. 7
CHUCKNOZZLE TIP
DRIVE AIR TUBE
DRIVE AIR
TUBE
NOZZLIE
TIP
SPRAY
LUBRICATION
CAN
EXHAUSTED
AIR TUBE
CHIP AIR
WATER