STERILIZATION
WARNING
Sterilize by autoclave after each patient use.
Sterilization is required before rst use.
Wear appropriate personal protective equipment during the sterilization and lubrication process. This should include eye protection and gloves.
Do not use ultrasonic bath, boiling water or chemicals to clean the handpiece.
Autoclave heating element could be located at the bottom of the sterilization chamber. To ensure that handpieces do not overheat, place
handpieces on central or upper tray.
Do not autoclave handpieces together with other instruments.
Do not use dry heat sterilization.
Keep handpiece in sterilization pouch until ready to use.
1. Wipe o handpiece with damp cloth.
2. Place the handpiece in a sterilization pouch and seal. Bag handpieces individually.
3. Steam autoclave for 15 minutes at 275
º
F (135
º
C ) in an autoclave.
4. Allow a minimum of 20 minutes for the drying cycle.