‘RETRY’ is displayed.
7-2 Error message RETRY appears.
The main causes when the error message RETRY appears:
When the image is not caught accurately because of the condition of the examinee (he/she
blinked or is not looking at the target etc.)
Alignment is not adjusted correctly
Failure of the main control board
Failure of the optical control board or CMOS camera
Failure of the measurement LED
YES
Check the condition of the eye to be measured.
Abnormal
Measurement cannot be taken in case
of disorder such as cataracts etc.
Abnormal
If the pupil is smaller than the reticle
mark, the measurement might not be
be taken.
Darken the surround and try to take a
measurement again.
Refer to 11-3
Normal
Realign correctly and take
a measurement again.
Refer to 6.2 Alignment of
the Operation Manual
Can a model eye be measured?
NO
Evaluate the Freeze Image.
Refer to 11