44
又は
Down
8.
Joystick Calibration
1 Move the cursor to 8. Joystick Calibration with
or
and press
<-|
.
2Make sure the cursor is positioned at R_LMT.
(The cursor is usually positioned at the place but if it is
not, move the cursor to the R_LMT* with UP or
DOWN switch.)
3Input the correction value of R indicated on the joystick ASSY.
R correction value
4Input the correction value of L, F, and B as same as 3) above.
5Press RETURN switch 2 times. The screen will go back to the measurement screen and
the detail of calibration (values) is stored.
If you turn the power off without pressing RETURN, the detail of calibration
will not be stored. Make sure to return to the measurement screen.
Up
** MAINTENANCE MODE **
1 Arr a n g e M o d e
2 F r ee z e I m a g e
3 S h i f t S P H D a t a
4 S h i f t T a r g e t D a t a
5 Pr i n t A d j u s t D a t a
6 H a r d w a r e C hec k
7 C a li b r a t i o n
8 Joystick C a li b r a t i o n
Up
Down
<-|
Return
PLC_T
107
FBC_T
107
R_LMT
82
L_LMT
-88
F_LMT
88
B_LMT
-94
Up
Down
()
(-)
Return