63
6) Open the program ComUpdate.
7) Select the COM Port (it depend on your PC, check it on Device Manager), Select
Baudrate to 115200 and Hyper to On.
8) Click on Select and choose the file SP9095_Vxxx.flash
9) Click on Send.
10) During update 3 green belt will appear on AKR750 screen.
11) When completed, press measuring button to exit from Program mode.
12) Switch off and switch on again AKR750 in Maintenance Mode to confirm the version of software.
13) Update of AKR750 is finished.