  • A
    Amy BarryAug 18, 2025
    What to do if Etac HIFLEX NEO shows AL84 error?
    • O
      omorrisonAug 18, 2025
      If your Etac Test Equipment displays error AL84, it indicates a power failure. Ensure the equipment is properly connected to a power source and that the power source is functioning correctly.
  • W
    Willie MaynardAug 20, 2025
    What to do if Etac HIFLEX NEO shows AL47 error?
    • D
      Dave AndersonAug 20, 2025
      If the Etac Test Equipment shows AL47, it means the temperature inside the chamber has exceeded the preset upper limit. Check your temperature settings and adjust them accordingly.
  • R
    Rebecca ColemanAug 23, 2025
    What to do if Etac Test Equipment shows AL49 error?
    • P
      Phillip ThomasAug 23, 2025
      If the Etac Test Equipment displays error AL49, it means the temperature inside the chamber has fallen below the preset lower limit value. You should check your temperature settings.
  • U
    ukochAug 26, 2025
    What to do if Etac HIFLEX NEO shows AL48 error?
    • J
      joseph07Aug 26, 2025
      If the Etac Test Equipment displays error AL48, it signifies that the humidity inside the chamber has exceeded the preset upper limit value. Review and adjust your humidity settings.
  • T
    Tony HartAug 28, 2025
    What to do if Etac Test Equipment shows AL50 error?
    • W
      Wendy DrakeAug 28, 2025
      If your Etac Test Equipment is showing error AL50, the humidity inside the chamber has dropped below the preset lower limit. Review and adjust your humidity settings.
  • E
    edwindeleonAug 30, 2025
    How to fix AL86 error on Etac HIFLEX NEO Test Equipment?
    • J
      james88Aug 30, 2025
      If your Etac Test Equipment displays error AL86, it's time to clean the refrigerator dust proof filter. Regularly cleaning this filter helps maintain optimal performance.
  • N
    Nicolas JohnsonSep 2, 2025
    What to do if Etac Test Equipment shows AL26 error?
    • J
      Jamie HerreraSep 2, 2025
      If your Etac Test Equipment is showing AL26, it means there is insufficient humidification water inside the water tank or subtank. Refill the water tank or subtank.
  • K
    Kathy MitchellSep 5, 2025
    Why does my Etac Test Equipment show AL73 error?
    • D
      dadkinsSep 5, 2025
      If your Etac Test Equipment shows AL73, it indicates insufficient humid water. Ensure that there is enough water for humidification.
  • T
    Thomas HallSep 7, 2025
    What does AL45 mean on Etac HIFLEX NEO?
    • R
      Richard BarrettSep 7, 2025
      If your Etac Test Equipment displays error AL45, the chamber ambient temperature has either fallen below 0 degrees or exceeded 40 degrees. Ensure the ambient temperature is within the specified range.
  • J
    Julie CooperSep 9, 2025
    What does AL52 error mean on Etac HIFLEX NEO Test Equipment?
    • M
      Martha EspinozaSep 9, 2025
      If your Etac Test Equipment displays error AL52, the ETACOM ambient temperature has exceeded 58 degrees. Ensure that the ambient temperature is within the allowed limits.

Summary

For Safe Use

Error Numbers

Alarm Numbers

How to Reset Error;Alarm

Error;Alarm Resetting Procedures

Step-by-step guide to reset active errors or alarms on the chamber.

Troubleshooting

When Er01 is Displayed

Detailed troubleshooting steps for the Er01 error, indicating chamber overheating.

When AL26 is Displayed

Troubleshooting for low humidification water level alarm.

Maintenance and Inspection

Local Setting (Control Panel)

Resetting an Error;Alarm

Procedures for clearing errors and alarms using the local control panel.

Remote Setting

Reset

How to reset errors or alarms remotely.

Troubleshooting Tips