2. Ensure the staple retaining cap is present during loading to minimize the potential
for premature sled movement, which could lead to inadvertent lockout.
Loading the device
ECHELON FLEX™ Powered & Powered Plus Staplers
Optimized Device Performance Guide
Proper reload selection: Tissue thickness should be carefully evaluated prior to using the instrument. Refer to the Reload Product Codes Chart, located in the Instructions for Use,
for proper reload selection. Selection of the appropriate staple reload should be based upon the combined thickness of both the tissue and the staple line reinforcement material.
4. Remove the staple retaining cap by lifting it straight up and discard.
5. Observe the surface of the new reload to ensure no drivers are visible and that the
reload lies flat in the jaw channel.
For complete product details, see Instructions for Use.
1. Position the instrument around the tissue to be stapled. When closing the device, ensure that the tissue lies flat and is positioned properly between the jaws. Any “bunching” of
tissue along the reload, particularly in the crotch of the jaws, may result in an incomplete staple line or inadvertent lockout. Holding the jaws in place for 15 seconds after closing
and prior to firing may result in better compression and staple formation.
Device closure
3. Insert the reload by sliding it against the bottom of the reload jaw until the
stapler alignment tab stops in the reload alignment slot of the jaw. Positioning
the reload at the bottom of the reload jaw could prevent inadvertent sled
movement and lockout.
1. Verify that the reload size matches the instrument size being used (e.g. use 60mm
reload with 60mm stapler).

Other manuals for Ethicon ECHELON FLEX

Question and Answer IconNeed help?

Do you have a question about the Ethicon ECHELON FLEX and is the answer not in the manual?

Ethicon ECHELON FLEX Specifications

General IconGeneral
BrandEthicon
ModelECHELON FLEX
CategoryStaple Gun
LanguageEnglish

Summary

Loading the Device

Verify Reload Size

Confirm reload size matches the instrument size for proper operation.

Ensure Staple Retaining Cap Presence

Verify the staple retaining cap is present to prevent premature sled movement and lockout.

Insert Reload into Jaw

Slide the reload into the jaw until the alignment tab engages the slot to prevent lockout.

Remove Staple Retaining Cap

Lift and discard the staple retaining cap after loading the reload.

Inspect Reload Surface

Check reload surface for visible drivers and ensure it lies flat in the jaw channel.

Device Closure

Position and Close Device

Ensure tissue lies flat between jaws during closure to prevent bunching and ensure proper staple formation.

Firing the Device

Fire Instrument and Return Knife

Pull firing trigger until motor stops; release to return knife home. Pulsing technique for compression is noted.

Stapler Lockout and Override

Use Knife Reverse Switch

Slide the knife reverse switch forward to attempt to return the knife to the home position.

Check Battery and Reinstall

Remove and reinstall the battery to ensure it is securely installed and has power.

Perform Manual Override

Use the manual override lever to return the knife blade to the home position if it fails to return.

Opening the Jaws

Ensure Knife is Home

Verify the knife blade indicator is in the home position under the cartridge jaw.

Release Jaws

Squeeze closing trigger, push anvil release button, then release trigger to open jaws.