EYELA A-1000S - User Manual

EYELA A-1000S
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Instruction
Manual
A-1000S
Tokyo Rikakikai Co., Ltd.
Aspirator
R00
This manual includes important information concerning the maintenance of
the functionality of the product and safety use.
Please keep this instruction manual within easy reach of yourself whenever using the product.
IMPORTANT
Please read this manual carefully especially on
Safety precautions before use.

Questions and Answers

  • J
    jason69Aug 6, 2025
    What to do if EYELA A-1000S motor doesn't work?
    • K
      Kenneth Valentine MDAug 6, 2025
      If the motor of your EYELA Medical Equipment isn't working even when the power switch is on, here are a few things to check: - Make sure the mains connector is properly plugged into the outlet. If it's loose, turn off the power and plug it in tightly. - Verify that power is being supplied by checking and turning on the breaker of the distribution board if necessary. - Inspect the fuse and replace it if it has blown. If the issue persists after replacing the fuse, stop using the equipment immediately and seek assistance from your local dealer or customer service center. - The thermal protector may be activated. Try lowering the ambient temperature or the water temperature of the bath.
  • F
    foleylauraAug 16, 2025
    How to fix poor vacuum degree in EYELA Medical Equipment?
    • J
      jenniferhaysAug 16, 2025
      If you're experiencing a poor vacuum degree with your EYELA Medical Equipment, consider these potential causes and solutions: - If solvent has dissolved into the water, replace the water in the aspirator bath, or prevent solvent from mixing into the water. - If the water temperature in the bath is too high, use the unit with Overflow, or place a cooling coil into the bath and connect it to a cooling water circulating device to lower the water temperature. - If the metal aspirator is clogged, remove the obstructing dirt. - Check the backflow prevention valve for clogs.
  • B
    brandismithAug 27, 2025
    Why is there poor air volume displacement in EYELA A-1000S?
    • P
      parsonskevinAug 27, 2025
      Poor air volume displacement in your EYELA Medical Equipment may be due to several factors: - Corrosion or wear of the metal aspirator. If this is the case, replace the metal aspirator. - Dirt clogged in the pump casing. Remove any dirt found in the casing. - Metal aspirator clogged with foreign substance. Remove the dirt. - If the pump part is cracked, stop operation immediately and contact your local dealer or customer service center. - Check the backflow prevention valve for clogs. - Water leaks from casing packing. - F cheese of pump part is damaged because of organic solvent.

Summary

Safety precautions

Signal word for warning

Defines warning signals and their meaning for user safety.

Check-up list of packing

Outline of the product

Use application

Describes the intended applications and usage of the aspirator.

Specification

Lists technical specifications of the Aspirator A-1000S model.

Characteristic of water vapor pressure

Presents a graph showing the relationship between water temperature and vapor pressure.

Attainable vacuum degree (reference)

Provides reference data for attainable vacuum degrees at different water temperatures.

Option

Details optional accessories like cooling coils and vacuum gauges.

Name of each part

Safety functions

Setup

Setup environment

Specifies the ideal environmental conditions for setting up the aspirator.

Utility connection

Details the necessary electrical and grounding connections for the aspirator.

Operation

Preparation

Outlines crucial preparation steps before operating the aspirator, including connecting hoses.

How to operate

Provides step-by-step instructions for operating the aspirator.

How to operate options

Explains how to use optional accessories like vacuum gauges.

How to operate options (cooling coil CCU and CCS)

Details the setup and operation of the cooling coil accessory.

Troubleshooting

Maintenance checkup

Disposal of the product

After-sale service

EYELA A-1000S Specifications

General IconGeneral
BrandEYELA
ModelA-1000S
CategoryMedical Equipment
LanguageEnglish