Note: If there is a serious odour problem, fail the sample at this
point of the test. Running the tester with seriously contaminated air
can damage the unit.
Press enter
Detector Tube Set-up
Note: CO and CO
2
are set. Water and oil channels are variable and
must be set as below:-
(v) Select water range by means of the up/down scroll buttons
HI 50 2000 mg/m³ for standard airline use
LO 5 200 mg/m³ for airlines incorporating a desiccant dryer.
Press enter
(vi) Select oil time
Screen will display:
Preset times
Add custom time
These are selectable by using the up/down scroll buttons
Press enter
Select according to the following requirements|:-
(vii) Preset times
The following common times are available within the preset times
menu and are selectable using the up/down scroll buttons 
Press enter
- 4 -