Note: For instructions on reading detector tubes and their safe
disposal, refer to the laminated sheet stored within the tester’s carry
case (also added as an appendix within this manual).
Refer to Appendix 4 within this manual for information converting
detector tube readings to a dewpoint.
(xiv) To restart the tester for a subsequent test, momentarily depress the
ON/OFF button. To switch off the tester, depress and hold the
ON/OFF button for approximately 2.5 seconds.
Note: The tester will automatically switch off, if not used, after a
period of 1 hour.
3.2 Testing high-pressure cylinders
(i) Turn on power by pressing and holding the ON/OFF button on the
keypad for approximately 2.5 seconds. The unit will display the
model and serial number.
Press enter
to confirm the date of the last calibration and the battery condition.
Note: It is recommended that your F3000 is calibrated within a 12-
month period.
(ii) If the battery condition is OK
Press enter
(iii) Ensure the system flow control valve is closed then connect a hose
from a regulated high-pressure supply to the air inlet on the tester.
A high-pressure regulator and hose assembly, reference F3002, is
available for this purpose. Ensure the cylinder is fully charged by
reading the content gauge. (A partly discharged cylinder can give
an incorrect water test result.)
Note: Maximum inlet pressure at the tester is 10 bar.
- 7 -