Press enter
IMPORTANT: If the unit is overpressurised it will automatically
shut down and display an overpressure warning; it will then
need to be returned to Factair for checking and resetting.
Note: The pressure displayed will be that of the regulated supply
from the cylinder, this must not exceed 10 bar.
(iv) Gently open system flow control valve until a minimal flow is
passing. Immediately carry out an odour test by smelling the air
exhausted from the sintered outlet above the flowmeter, then close
the system flow control valve. Use the minimum amount of air to
preserve cylinder capacity.
Note: If there is a serious odour problem, fail the sample at this
point of the test. Running the tester with seriously contaminated air
can damage the unit.
Press enter
Detector Tube Set-up
Note: CO and CO
2
are set. Water and oil channels are variable and
must be set as below:-
(v) Select ‘LO 5 200 mg/m³ water range by means of the up/down
scroll buttons 
Press enter
(vi) Select oil time
Screen will display:
Preset times
Add custom time
These are selectable by using the up/down scroll buttons
Press enter
Select according to the following requirements:-
- 8 -