TRACHEOPORT
®
JUNIOR
GEBRAUCHSANWEISUNG • INSTRUCTIONS FOR USE

Table of Contents

Questions and Answers

  • J
    Jonathan ButlerAug 12, 2025
    What to do if my Fahl TRACHEOPORT JUNIOR Medical Equipment device fails to start?
    • W
      William FisherAug 12, 2025
      If your Fahl Medical Equipment device isn't starting, there might be a few reasons. First, the rechargeable battery could be discharged; connect the power supply and charge the battery for 1-2 hours. Second, the mains plug of the power supply might not be well seated; check all plug connections and ensure the indicator light is lit. Finally, there might be a vacuum in the container; de-aerate the container.
  • A
    Alyssa BrownAug 18, 2025
    Why is my Fahl TRACHEOPORT JUNIOR Medical Equipment showing weak or no vacuum?
    • J
      James FritzAug 19, 2025
      A weak or absent vacuum in your Fahl Medical Equipment could stem from several causes. Check if the bacterial filter is missing and insert it if necessary. Inspect the tubing and secretion container lid for leaks, ensuring all connections are secure. It's also possible that secretions or blood have been aspirated into the device, causing the valve platelets to stick; in this case, verify the assembly position of the secretion container lid.
  • B
    Brian LesterAug 27, 2025
    What to do if my Fahl TRACHEOPORT JUNIOR has insufficient power?
    • K
      Kyle TorresAug 27, 2025
      If your Fahl Medical Equipment is showing insufficient power, the rechargeable battery may be discharged; charge the battery. Alternatively, the filter could be obstructed, in which case you should replace it. Another potential cause is a leak in the tubing or the secretion container lid; check these for firm seating.
  • L
    laura88Sep 3, 2025
    Why is my Fahl TRACHEOPORT JUNIOR showing strong vacuum?
    • M
      Matthew SmithSep 3, 2025
      A strong vacuum indication on your Fahl Medical Equipment may be due to an obstructed bacterial filter; replace the bacterial filter. Alternatively, the float of the overflow protector might be blocking the secretion container lid inlet; check the inlet, and if necessary, empty the container, clean the overflow protector, and ensure the float moves freely.

Fahl TRACHEOPORT JUNIOR Specifications

General IconGeneral
Product NameFahl TRACHEOPORT JUNIOR
CategoryMedical Equipment
TypeTracheostomy Tube
SterilitySterile
Single UseYes
PackagingIndividually packaged
MaterialPVC