A-18 ENGLISH
Technical specifications
PRODUCT SPECIFICATIONS
ADULT NEONATAL
SPECIFICATION VALUE VALUE
Dimensions
(heaterbase only)
240 mm (D) x 154 mm
(W) x 253 mm (H)
240 mm (D) x 154 mm
(W) x 253 mm (H)
Weight (heaterbase
and power cord only)
3�45 kg 3�45 kg
Supply frequency 50/60 Hz 50/60 Hz
Supply voltage
950AXX
1
230V
950JXX
1
115V
950GXX
1
100V
950AXX
1
230V
950JXX
1
115V
950GXX
1
100V
Power rating 350 VA 350 VA
Sound pressure level Alarms exceed 45 dBA
@ 1 m
Alarms exceed 45 dBA
@ 1 m
Auditory alarm pause 120 seconds 120 seconds
Humidity performance
(except in the event
of a humidifier alarm
or power failure or
electromagnetic
disturbance)
Invasive mode (37 °C):
>33 mg/L
Mask mode (31 °C):
>12 mg/L
Optiflow mode (37 °C):
>12 mg/L
Neonatal mode (37 °C):
>33 mg/L
Maximum temperature
of delivered gas
43 °C 43 °C
Operating flow
range (at 22 °C room
temperature)
Invasive mode:
5 – 60 L/min
Mask mode:
5 – 120 L/min
Optiflow mode:
5 – 70 L/min
Neonatal mode:
0�5 – 40 L/min
See breathing circuit user
instructions for specific
flow ranges
Time to reach set
temperature (gas flow
is required)
<30 minutes <30 minutes
Maximum surface
temperature of the
breathing circuit
(applied part section)
44 °C 44 °C
Temperature
measurement accuracy
± 2 °C ± 2 °C
Component service life Heaterbase: 7 years Heaterbase: 7 years
1
XX represents the country code