A-19 ENGLISH
Technical specifications
OPERATING CONDITIONS
ADULT NEONATAL
SPECIFICATION VALUE VALUE
Room temperature 18 – 26 °C 20 – 26 °C
Incoming gas
temperature
Minimum = Room
temperature
Maximum = 10 °C above
room temperature (at
30% relative humidity)
Minimum = Room
temperature
Maximum = 10 °C above
room temperature (at
30% relative humidity)
Operator position <1 m from heaterbase <1 m from heaterbase
Atmospheric pressure Minimum of 70 kPa
(equivalent to a
maximum altitude of
3000 m)
Minimum of 70 kPa
(equivalent to a
maximum altitude of
3000 m)
STORAGE CONDITIONS
ADULT NEONATAL
SPECIFICATION VALUE VALUE
Temperature -20 – 60 °C -20 – 60 °C
Humidity 10% – 95%
relative humidity
non-condensing
10% – 95%
relative humidity
non-condensing
Dispose of according to standard hospital procedure and national regulations for
electrical and electronic equipment�
Note that a lithium cell battery is included in the sensor cartridge