USER MANUAL

Table of Contents

Questions and Answers

  • D
    Danielle BurtonSep 15, 2025
    What to do if the pulse ox sensor is off the patient when using F&P Medical Equipment?
    • H
      Helen GibsonSep 16, 2025
      Check that the sensor is properly attached to a suitable measurement site on the patient when using F&P Medical Equipment.
  • J
    joseduffyAug 25, 2025
    What to do if the F&P Medical Equipment detects a blockage?
    • K
      Kelsey Mcgrath DDSAug 25, 2025
      If the Airvo 3 Medical Equipment detects a blockage, check the heated breathing tube, patient interface, and inlet air filter for obstructions. Also, ensure the patient interface is the correct size and the target flow rate is within range. If problems persist, replace the consumables.
  • J
    Jeffrey WillisAug 26, 2025
    How to fix a device fault on F&P Medical Equipment?
    • J
      John LittleAug 26, 2025
      To resolve a device fault with your F&P Medical Equipment, restart the device. If the issue persists after restarting, contact your service representative.
  • D
    duncandanaAug 26, 2025
    What to do if F&P Medical Equipment shows 'Check Tube'?
    • J
      Justin LindseyAug 27, 2025
      If the Airvo 3 Medical Equipment cannot detect the heated breathing tube, check that the tube is not damaged and is plugged in correctly. If the problem continues, replace the heated breathing tube.
  • S
    Sylvia DeanSep 16, 2025
    What to do if disinfection failed to hold temperature on F&P Medical Equipment?
    • M
      moralesnicoleSep 16, 2025
      Check that the disinfection tube blue connector is connected to the top of the outlet elbow, the disinfection tube red end is connected to the left hand chamber port, the disinfection filter is connected to the right hand chamber port. Then restart the device. If the problem is not resolved, replace the disinfection tube and outlet elbow in turn on your F&P Airvo 3 Medical Equipment.
  • M
    Matthew WashingtonSep 23, 2025
    What to do if wall power is disconnected during disinfection on F&P Medical Equipment?
    • J
      Jonathan AllenSep 23, 2025
      Connect device to wall power, then restart the device on your F&P Airvo 3 Medical Equipment.
  • K
    Kelly BartonSep 8, 2025
    What to do if F&P Airvo 3 Medical Equipment indicates 'Critically Low Battery'?
    • O
      ocaldwellSep 8, 2025
      If the Airvo 3 Medical Equipment indicates a critically low battery, connect it to a wall power supply immediately to maintain therapy. Humidification is turned off to maintain operation of the blower and oxygen supply.
  • J
    Julia RothSep 9, 2025
    What to do if there's no O2 pressure at the HPO port on F&P Medical Equipment?
    • J
      Jonathan RhodesSep 9, 2025
      If there is no oxygen being supplied to the high-pressure (HPO) inlet port during therapy with your F&P Airvo 3 Medical Equipment, check that the oxygen supply is working. If using an oxygen bottle, check the bottle is not empty. If switching to the low-pressure (LPO) inlet port or stopping oxygen delivery, set the FiO2 target to 21 %.
  • D
    Debra PatelSep 2, 2025
    What to do if the 'Outlet Elbow' is too warm on F&P Medical Equipment?
    • N
      Natalie AndersonSep 2, 2025
      If the outlet elbow is too warm to run startup checks on your F&P Airvo 3 Medical Equipment, wait for the outlet elbow to cool down. If the issue persists, replace the outlet elbow.
  • S
    Sarah NashAug 31, 2025
    Why is the flow above target on F&P Medical Equipment?
    • T
      taramezaAug 31, 2025
      If the Airvo 3 Medical Equipment flow rate is higher than the target, check for leaks in the water chamber, heated breathing tube, and patient interface. Ensure the inlet air filter is correctly inserted and the target flow rate is within range. If the issue persists, replace the consumables.

Summary

Getting Started

Additional Resources

Information on where to find more help and resources for using the Airvo 3 device.

Manual Conventions

Explains the meaning of warning, caution, and note symbols used throughout the manual.

Introduction and Safety

Airvo 3 Introduction

Overview of the Airvo 3, its intended use, indications, contraindications, and side-effects.

Essential Safety Precautions

Covers general warnings, oxygen safety, and pulse oximetry safety precautions.

Device Overview, Setup, and Operation

Device Components and Interface

Identifies system and device components, and explains user interface navigation.

Airvo 3 Setup and Preparation

Steps for setting up the device, equipment, and oxygen connections.

Using the Airvo 3 for Therapy

Details on therapy settings, starting, and managing operation.

Monitoring, Troubleshooting, and Reprocessing

Monitoring Patient Data

Accessing and interpreting patient data, live values, and long-term graphs.

Troubleshooting Alarms and Issues

Guidance on resolving common alarms, faults, and device problems.

Device Reprocessing Procedures

Instructions for cleaning and disinfecting the Airvo 3 and its accessories.

Pulse Oximetry and Specifications

Pulse Oximetry Setup and Use

Connecting, configuring, and using pulse oximetry for patient monitoring.

Technical Specifications

Provides detailed technical data for the Airvo 3 device and its operations.

F&P Airvo 3 Specifications

General IconGeneral
BrandF&P
ModelAirvo 3
CategoryMedical Equipment
LanguageEnglish