6 ENGLISH TECHNICAL INFORMATION
OXYGEN/PRESSURE PORT CONNECTOR
If pressure readings and/or additional oxygen are required, an Oxygen/Pressure Port Connector is
available (REF 900HC452). Please contact your healthcare provider.
TECHNICAL SPECIFICATIONS
The F&PVitera Full Face mask has exhaust holes to expel the air that you exhale from the mask. It is
important that these exhaust holes are not blocked by any object. This controlled leak ensures all exhaled
CO
2
is expelled from the mask.
F&PVITERA EXHAUST FLOW INFORMATION
Pressure (cmH
2
O) 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Flow (L/min) 24 28 32 36 40 43 47 50 53 55 58 60 62 64
Due to manufacturing variations, the exhaust flow rates may vary from the nominal values shown above.
Exhaust Flow (L/min)
Pressure (cmH
2
O)
60
50
30
10
70
40
20
0
4 5 7 9 11 13 15 17 19 21 23
25
RESISTANCE TO FLOW
Pressure drop through mask at 50 L/min: 0.24±0.15 cmH
2
O
Pressure drop through mask at 100 L/min: 0.47±0.15 cmH
2
O
INSPIRATORY AND EXPIRATORY RESISTANCE
Inspiratory resistance of the mask (in combination with the Non-Rebreathing Valve) is 1.05±0.15 cmH
2
O
Expiratory resistance of the mask (in combination with the Non-Rebreathing Valve) is 0.80±0.15 cmH
2
O
The opening to atmosphere pressure of the Non-Rebreathing Valve is <0.56 cmH
2
O
The closing to atmosphere pressure of the Non-Rebreathing Valve is <2.16 cmH
2
O
MASK DEAD SPACE
Small = 246 cc Medium = 271 cc Large = 322 cc
SOUND
The sound power level of mask is 29.8 dBA, with uncertainty 2.5 dBA
The sound pressure level of the mask is 21.8 dBA, with uncertainty 2.5 dBA
ADDITIONAL INFORMATION
The F&PVitera Full Face mask and accessories are not made with natural rubber latex or DEHP.
For single patient reuse, the mask should be replaced no later than 12 months after the date of first use.
Store dry mask in clean conditions out of direct sunlight. Storage temperature: -20 to 50°C (-4 to 122°F)
MULTIPATIENT REPROCESSING INSTRUCTIONS
Your F&PVitera Full Face mask is provided patient-ready.
To reuse between patients please follow the guidelines below or visit www.fphcare.com/disinfection.
If you do not have internet access, please contact your F&P Healthcare representative.
At point of use, remove any gross soiling on the mask by wiping or rinsing in potable water.