Model 35 Viscometer
Instruction Manual
Manual No. 208878, Revision N
Models 35A, 35SA, 35A/SR-12, and 35SA/SR-12

Table of Contents

Questions and Answers

  • A
    Angel RoweAug 9, 2025
    Why are my Fann Measuring Instruments out of calibration?
    • T
      Tara DyerAug 9, 2025
      If your Fann Measuring Instruments are out of calibration, it could be due to several reasons. Contaminated bob shaft bearings can cause this, and the solution is to replace the bob shaft bearings (P/N 207450). A bent bob shaft can also be the culprit; try to slightly bend the shaft to straighten it, or contact Fann for repair or replacement. Other possible causes include a bent rotor (replace the rotor), a damaged or incorrectly installed torsion spring (refer to Section 5.4 to replace the torsion spring), or an incorrect motor speed (replace the motor with P/N 207446 or 207447).
  • S
    Shane FloresAug 16, 2025
    What to do if my Fann Measuring Instruments have erratic dial motion?
    • M
      Mrs. Nancy StevensAug 17, 2025
      Erratic dial motion in Fann Measuring Instruments can stem from a few issues. One potential cause is contaminated bob shaft bearings, which can be resolved by replacing the bob shaft bearings (P/N 207450). Another cause might be a bent bob shaft; in this case, try bending the shaft slightly to straighten it, or contact Fann for repair or replacement. Finally, a rotor out of alignment or damage to the rotor can also lead to erratic dial motion, so replace the rotor if necessary (refer to Table 9-1 for a list of rotors).
  • T
    Timothy QuinnAug 25, 2025
    What causes excessive run-out of rotor in Fann Measuring Instruments?
    • C
      Christopher WhiteAug 25, 2025
      Excessive run-out of the rotor in Fann Measuring Instruments can be caused by a damaged rotor, which requires replacement (refer to Table 9-1). Another potential cause is contamination in the main shaft recess, in which case you should contact Fann for repair.
  • D
    Daniel RandolphSep 1, 2025
    How to fix excessive noise in Fann Measuring Instruments?
    • T
      Tanya GravesSep 1, 2025
      Excessive noise in Fann Measuring Instruments may be due to lubrication failure or contamination in the gears, in which case you should contact Fann for repair. It could also be caused by a worn center thrust washer, which also requires you to contact Fann for repair. Another possible cause is that the top cover is not set properly; try adjusting the top cover.

Summary

Safety Procedures

Safe Electrical Operation

Provides guidelines for safe electrical operation, including handling power and avoiding hazards.

Viscometer Operation Procedures

Operating Six-Speed Models (35 A;35 SA)

Details how to operate the six-speed models using speed and gear shift controls.

Operating Twelve-Speed Models (35 A;SR-12;35 SA;SR-12)

Explains the operation of twelve-speed models with an additional gear box lever.

Instrument Calibration

Test Analysis and Calculations

Newtonian Viscosity Calculation

Explains how to calculate Newtonian viscosity using dial readings and factors.

Plastic Viscosity and Yield Point

Details the method for calculating Plastic Viscosity (PV) and Yield Point (YP).

Troubleshooting and Maintenance

Troubleshooting Common Issues

Offers a guide to identify and correct common problems encountered with the viscometer.

fann 35SA Specifications

General IconGeneral
Brandfann
Model35SA
CategoryMeasuring Instruments
LanguageEnglish