Liability for damage from product use transferred to owner/operator if used improperly or by unqualified persons.
Holding device for fasciotens products during surgical procedures; class I medical device.
Combination with other fasciotens products or operating tables/standard rails.
Use is contraindicated if a sufficiently stable fastening rail is not available.
Reusable medical device; service life depends on wear and damage; processing does not affect performance.
Disassemble instruments, loosen eccentric locks before cleaning and transport in a sealed container.
Post can be disassembled into individual parts for processing, identified by serial number with 'Z'.
Consists of manual, ultrasonic, or machine cleaning steps using alkaline cleaners.
Instruments can be sterilised individually wrapped using steam with fractionated pre-vacuum at 134°C.
Processor ensures validation and routine monitoring for desired processing results.
Store in a clean, cool, dry location, protected from damage and handled with care.
Regular inspections and maintenance by fasciotens GmbH ensure functionality and safety.
Contact customer support for issues; all repairs must be carried out by the manufacturer.
Legal warranty period of 24 months; initial defects must be reported to Customer Support.
Contact Customer Support team via email or telephone for any issues or questions.
| Product Name | fasciotens Carrier HC010 |
|---|---|
| Category | Medical Equipment |
| Sterility | Sterile |
| Single Use | Yes |
| Material | Silicone |
| Application | Abdominal wall closure |
| Indication | Temporary abdominal closure |
| Compatibility | Compatible with negative pressure wound therapy |