FEASA LED ANALYSER
FUNCTIONAL VERSION
Accuracy White X=± 0.01, Y = ± 0.01 @ x=0.33, y=0.33 using OH-3 Optical Head
Red (630nm) ± 4nm
Green (520nm) ± 2nm
Blue (450nm) ± 5nm
Yellow (590nm) ± 2nm
These accuracy's are based on the Led Analyser being calibrated at a point in time against known set values.
Capture Test Times Auto Range 350ms
Range 1 ( Low Intensity) 650ms
Range 2 ( Medium Intensity) 200ms
Range 3 ( High Intensity) 22ms
Range 4 ( Super Intensity) 4ms
Range 5 ( Ultra Intensity) 2ms