Medication chamber is loosened and
not completely attached.
Connection of AC adapter to nebulizer
is loosened.
Medication has run out.
No contact between medication and
vibrating mesh for more than 10
seconds.
Nebulizer is being shaken in the use.
Re-attach the medication chamber correctly
and restart the power.
Re-connect in the correct manner and
restart the power.
Put in the medication in the medication
chamber.
Adjust the nebulizer’s angle so the
medication can come in contact.
Hold the nebulizer in the hand stably.
Medication chamber or vibrating mesh
is broken.
Replace with a new medication chamber
and then put in the medication.
Nebulizer
shuts off in
usage.
Overflow of
medication
from
medication
cup chamber
Rupture of medication chamber or
aging of silicone ring.
Replace with a new medication chamber
and then put in the medication.
If your nebulizer still does not function properly after taking the solution mentioned above,
please contact the retailer from which you purchased the product.
11. Note & Warning
11.1 Note and suggestion
• Nebulizer is a medical device,please read the user manual before using it.
• Please use required accessories, warranty service is not provided for damage caused by accessory
beyond our list.