Guidance and manufacturer ’s declaration-electromagnetic immunity
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
Mains power quality should
be that of a typical commercial
or hospital environment.
±2 kV for a.c.
power lines
±1 kV for d.c.
power lines
N/A
N/A
Mains power quality should
be that of a typical
commercial or hospital
environment.
±2, ±4, ±6kV
for Contact
discharge
±2,±4,±8kV,
air discharge
±2, ±4, ±6kV for
Contact discharge
±2,±4,±8kV, air
discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at
least 30%.
IEC60601
test level
Compliance level
Electromagnetic
environment -guidance
The infrared thermometer is intended for use in the electromagnetic environment
specified below. The customer or the user of the infrared thermometer should assure
that it is used in such an environment.
Guidance and manufacturer ’s declaration-electromagnetic emissions
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class B
N/A
N/A
The infrared thermometer uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause and interference in
nearby electronic equipment.
The infrared thermometer is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply.
network that supplies buildings used for
domestic purposes.
Compliance Electromagnetic environment-guidance
The infrared thermometer is intended for using in the electromagnetic environment
specified below. The customer or the user of the infrared thermometer should assure
that it is used in such an environment.
21. EMC information