Temperature Settings | 31°C, 34°C, 37°C |
---|---|
Power Supply | 100-240V AC, 50/60Hz |
Therapy Modes | High Flow Nasal Cannula (HFNC) |
Humidity Output | 44 mg/L |
FiO2 Range | 21-100% |
Operating Humidity | 10% to 95% RH, non-condensing |
Flow Rates | 2-60 L/min |
Specifies the intended purpose and patient population for the AIRVO 2 device.
Lists critical safety precautions related to burns and electric shock.
Covers initial placement, unboxing, and attaching the water chamber with safety warnings.
Instructions for connecting the sterile water bag for automatic chamber filling.
Details on installing the breathing tube and associated safety precautions.
Steps for powering the unit on and verifying its disinfection status.
Explains the warm-up process and how to activate Junior Mode.
Details on setting the target dew-point temperature, including available levels and limitations.
Instructions for setting the target flow rate, including available ranges and limitations.
Explanation of how to activate and deactivate the transport mode for specific use cases.
Critical warnings regarding the safe use of oxygen with the AIRVO 2 unit.
Steps for connecting oxygen and adjusting the level, including display interpretation.
Guidance on operating the device during use and powering it off afterward.
Describes alarm signals and details alarm messages, causes, solutions, and delivery impact.
Lists alarm limits and procedures to test the alarm system functionality.
Table outlining the recommended replacement intervals for various accessories.
Steps for filter replacement and notes on servicing.
Lists detailed technical specifications of the AIRVO 2 device.
Specifies operating, storage, and transport environmental conditions.
Provides guidance on the proper disposal of the device and its consumables.